Product Description

Automated nucleic acid workstation


 

Product overview

Designed with rotary nucleic acid extraction technology, PANA9600S is the latest generation of automated nucleic acid workstation independently developed by Tianlong, with the support by the National Special Fund for Major Scientific Instrument Development (No. 2012YQ030261). The workstation integrates sample loading, nucleic acid extraction, PCR system construction and many other functions in one, and can achieve the automated operation of the whole process. Equipped with high performance of the specific nucleic acid extraction kits, this instrument can extract the required nucleic acids from a wide range of samples such as whole blood, serum/plasma, swab solution, urine, etc. in a shorter time to ensure the specific downstream applications.

Features:

Reliable pollution control ensured by rotary nucleic acid extraction technology used with droplet capture, experimental partitioning, air filtration, UV disinfection and other measures.

Fast and high throughput: automated loading, nucleic acid extraction and PCR system setup for 96 samples within 70min.

Highly flexible and adaptable: can be applied with single-strip kits, 3- to 6-well extraction available to process various sample types; 4 PCR reaction systems can be set up simultaneously.  

Accurate and reliable: precise sample loading, accurate temperature control, provide stable and reliable test results with reagents and consumables independently developed by Tianlong.

Human-centered design, automated operation:  pre-encapsulated extraction kits, and automated preparation of the PCR systems. Real whole-process automation with automatic dilution of proteinase K, and automatic sample retention of the original sample and the nucleic acid extracted, among other easy-to-use design.

Product Model

PANA9600S

Authoritative Certifications:

NMPA, CE, KFDA, etc.

Sample Throughput

1-96

Features

Automatic scanning and identification of sample information + nucleic acid extraction + PCR system setup

Principle

The principle of magnetic bead method, and the technology of rotational mixing

Detectability

192 assays can be completed per run, with 4 different systems of PCR-setup simultaneously.

Specimen Types

Samples of plasma, serum, whole blood, swab solution, and urine, etc.

Sample Loading Channels

1-4

Pipetting Performance

15μL and below: accuracy: A≤2%, repetitive CV ≤ 3.0%;

15-50μL: accuracy: A≤1.5%, repetitive CV≤1.5%;
50μL and above: accuracy: A≤1.0%, repetitive CV≤1.2%;

Liquid Level Detection

Capacitive-sensing level detection; pressure-sensing level detection

Sample Tube Specifications

Compatible with a wide range of sample collection tubes, 1.5mL and 2.0mL centrifuge tubes, cryopreservation tubes and sample cups.

Extraction Temperature Control Module

2 steps: pyrolysis & elution,

with temperature control from 35 to 120 °C.

Visual Inspection Technology

Real-time inventory status monitoring of consumables and reagents

PCR Reagent Chamber

Light-proofing design; automatic cooling on start-up (4 ℃ ~ 15 ℃ refrigerator)

PCR Supplies/Consumables

Compatible with 0.1mL and 0.2mL 8-tube strips, and 96-well plates;

Contamination Control Design

(1) Separate, enclosed nucleic acid extraction space, with directional top ventilation to create an internal negative pressure system;

(2) Sample loading device equipped with airtight, drip-proof design.

(3) External droplet capture tray;

(4) UV light sterilization units for the experimental and extraction cabinets;

(5) Customized function: bottom ventilation of the nucleic acid extraction area available by special ventilation circuit design.

Information technology

(1) Multiple sample barcodes are scanned one by one during sample rack loading.

(2) Identification of the extraction kit barcode, and automatic parsing of the extraction program information.

(4) Identification of PCR kit component barcode, and automatic parsing of amplification program information.
(5) Automatic parsing of PCR information and LIS connection, based on nucleic acid extraction data.

Operation

Embedded 12-inch touch screen; multi-module real-time status light

Size and Weight

1370mm x 810mm x 890mm (W*D*H), 220 kg

Power

AC 220± 20 V, 50/60 Hz.